3.1.17. Submenu “Diagnostics”
This function is required to carry out a basic test of the load cell, the instrument and their
interconnections.
3.1.17.1. ADC – Integer
output “I - INT”
This function shows the direct output
value of the 24-Bit-ADC.
To read this integer value in the range
of 0 … 4194303 is the basic requirement
to be ready for calibration.
0
= 0 mV/V
2000000
= 2 mV/V
3.1.17.2. ADC – input ratio
mv/V “I – MV”
This output is a floating-point result
of the current input voltage ratio in
mV/V checking the load cell
arrangement.
3.1.17.3. ADC – output
normalized “I – nOM
This function can only be used after
calibration has been carried out.
It shows the current weight on the load
cell as percentage of max value.
3.1.17.4. ADC – ouput at zero
“I – ZE”
This function can only be used after
calibration has been carried out. It
gives the input voltage ratio in mV/V
at calibrated zero. It is recommended
to note it down for further use.
3.1.17.5. ADC – output “I –
MX”
This function can only be used after
calibration has been carried out. It
gives the input voltage ratio in mV/V
at calibrated max value.