3.1.9. Submenu “Print”
This menu defines the major parameter to get a direct communication to a printer.
3.1.9.1. Printer enable “EN”
This toggled decision defines whether the
printer port is enabled for further use.
default setting: on
3.1.9.2. Printer port select “PoRT”
Set in terms of numbers to defines the
physical address of the communication
port at A810 to the printer.
default setting: 2 (Pins E19, 20)
Refer to 2.2 at page 12.
3.1.9.3. Baudrate select “Bd”
Set the desired baud rate for the
communication port of A810 to the
printer in the range of
1200 … 76800 baud.
default setting: 9600
3.1.9.4. Parity check “PAR”
Select the number of data bits for
transmission and the type of parity.
default setting: 8no
8no
8 databits, no parity, 1stopbit
8EvE
8 databits, even parity,
1stopbit
8odd
8 databits, odd parity, 1stopbit
7EvE
7 databits, even parity,
1stopbit
7odd
7 databits, odd parity, 1stopbit
E N
o F F E N
o n E N
/NET
GROSS
0
0
TARE
/NET
GROSS
ENT
ENT
P o R T
2
/NET
GROSS
0
0
/NET
GROSS
ENT
-1
+1
value
incr. decr.
B d
9 6 0 0
B d
/NET
GROSS
0
0
TARE
/NET
GROSS
ENT
-1
+1
value
incr. decr.
P A R
8 n o
P A R
/NET
GROSS
0
0
TARE
/NET
GROSS
ENT
-1
+1
value
incr. decr.
TARE
TARE
TARE
TARE