3
Overview
12
STERRAD VELOCITY Reader User’s Guide
System Status Indicator
The System Status indicator light indicates when the device is powered on by
displaying a blue light. When scanning the BI barcode, the blue light flashes brighter to
indicate a successful scan. If the system detects an error the status indicator changes
to red. The system status indicator will blink red when a system status error is detected,
and it will remain solid red when a fatal error is detected. The indicator would display
a red light for issues such as: the device is unable to read from the internal storage,
the internal clock returns an erroneous reading, or the device detects a voltage or
temperature outside of a specific range. Refer to the “Troubleshooting” section of this
guide (Chapter 8) if a red system status indicator is displayed.
System Status Indicator Light
Indicator State
Meaning
Solid Blue
System Status: OK
Bright Blue Flash
Visual affirmation that biological indicator
barcode has been scanned successfully.
Blinking Red
System Status: System Failure
Solid Red
System Status: General System Error
Table 1.
System Status Indicator
Biological Indicators
The BI is a single-use testing device used to monitor the effectiveness of
hydrogen peroxide sterilization. STERRAD VELOCITY Biological Indicator / Process
Challenge Devices are for use in STERRAD Sterilizers only. The BI is designed to monitor
all cycles of the STERRAD NX with and without ALLClear Technology, STERRAD 100NX
with and without ALLClear Technology, and STERRAD 100S Sterilization Systems.
Each biological indicator includes a chemical indicator on the top of the cap which
changes color from red/pink to yellow, or yellow with some red/orange/brown dots,
when exposed to hydrogen peroxide (refer to the “Chemical Indicator Result” section
(Chapter 4) for more detailed instructions). During the sterilization cycle hydrogen
peroxide passes into the interior of the BI through the sterilant ingress windows and
kills the indicator organism. Once the sterilization cycle is complete the BI is placed
in one of eight biological indicator wells, on the top of the reader, for incubation and
reading. The BI wells detect the presence of the BI and begin the incubation and
interrogation process.
Содержание Sterrad Velocity Reader
Страница 1: ...User s Guide AB11321031 AB11321031 50 January 2021 ASP 2021 All rights reserved Reader...
Страница 6: ...vi STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 14: ...8 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 36: ...30 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 40: ...34 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 52: ...46 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 66: ...THIS PAGE INTENTIONALLY LEFT BLANK 60 STERRAD VELOCITY Reader User s Guide...
Страница 68: ...THIS PAGE INTENTIONALLY LEFT BLANK 62 STERRAD VELOCITY Reader User s Guide...
Страница 70: ...THIS PAGE INTENTIONALLY LEFT BLANK 64 STERRAD VELOCITY Reader User s Guide...
Страница 72: ...THIS PAGE INTENTIONALLY LEFT BLANK 66 STERRAD VELOCITY Reader User s Guide...