4
Operation
24
STERRAD VELOCITY Reader User’s Guide
Negative Test Result
A green well number indicates a negative result for a test BI. Touch the green well
indicator number or remove the BI to go to Result Confirmation. In Result Confirmation,
view the result summary by touching
View and Print Result Summary
. Touch
OK
or
remove the test BI from the well to return to the Status Screen. Touching
View and Print
Result Summary
opens the Result Summary page. If a printer is connected to the reader,
touch
on the lower left to print the Result Summary. Touch
Add Note
to leave notes
about this specific result. Touch
Done
or remove the BI to return to Result Confirmation.
Figure 18.
Negative Test Result
Positive Test Result
A positive result for a test BI will be accompanied by an audible alarm and a red
warning screen which indicates the well where the positive result occurred. Touch
Continue
or remove the BI to silence the alarm and go to the next steps. Enter the operator’s
name and follow the directions for handling a positive result. Touch
Done
when finished.
9
Note:
If a positive result is observed:
1. Follow the current hospital or healthcare facility policies and procedures
regarding quarantine, or retrieval, and reprocessing of potentially non-sterile
instruments and notification of the clinician(s).
2. Repeat the test with a second BI.
Содержание Sterrad Velocity Reader
Страница 1: ...User s Guide AB11321031 AB11321031 50 January 2021 ASP 2021 All rights reserved Reader...
Страница 6: ...vi STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 14: ...8 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 36: ...30 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 40: ...34 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 52: ...46 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 66: ...THIS PAGE INTENTIONALLY LEFT BLANK 60 STERRAD VELOCITY Reader User s Guide...
Страница 68: ...THIS PAGE INTENTIONALLY LEFT BLANK 62 STERRAD VELOCITY Reader User s Guide...
Страница 70: ...THIS PAGE INTENTIONALLY LEFT BLANK 64 STERRAD VELOCITY Reader User s Guide...
Страница 72: ...THIS PAGE INTENTIONALLY LEFT BLANK 66 STERRAD VELOCITY Reader User s Guide...