8
Troubleshooting
54
STERRAD VELOCITY Reader User’s Guide
Problem/Error Message
Possible Cause
Suggested Solution
Test BI positive result.
Sterilization not
achieved. May be
due to overloaded
chamber, wet
load, material
incompatibility,
BI misuse, or
defective BI.
Follow the current hospital
or healthcare facility policy
and procedures regarding
quarantine, or retrieval, and
reprocessing of potentially
non-sterile instruments and
notification of the clinician(s).
Repeat the test with a second
BI. Check BI for any damage per
IFU prior to use for cycle. Ensure
the BI is handled and activated
properly. Do not place tape or
labels on the BI cap or growth
reservoir. If problem persists,
contact ASP.
First Test BI positive
result second test BI
negative result.
Fluorescent
contamination on
BI, sterilization not
achieved, improperly
placed BI, ingress
hole blocked. May
be due to overloaded
chamber, wet
load, material
incompatibility,
BI misuse, or
defective BI.
Test a third BI to confirm
sterilizer performance.
Содержание Sterrad Velocity Reader
Страница 1: ...User s Guide AB11321031 AB11321031 50 January 2021 ASP 2021 All rights reserved Reader...
Страница 6: ...vi STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 14: ...8 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 36: ...30 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 40: ...34 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 52: ...46 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 66: ...THIS PAGE INTENTIONALLY LEFT BLANK 60 STERRAD VELOCITY Reader User s Guide...
Страница 68: ...THIS PAGE INTENTIONALLY LEFT BLANK 62 STERRAD VELOCITY Reader User s Guide...
Страница 70: ...THIS PAGE INTENTIONALLY LEFT BLANK 64 STERRAD VELOCITY Reader User s Guide...
Страница 72: ...THIS PAGE INTENTIONALLY LEFT BLANK 66 STERRAD VELOCITY Reader User s Guide...