Operation
4
STERRAD VELOCITY Reader User’s Guide
17
If it is customary for your department to confirm chemical indicator color change
on a positive control BI, touch
Red / Pink
if the chemical indicator on the BI cap has
remained red/pink (refer to the “Chemical Indicator Result” section (Chapter 4) for more
detailed instructions). Touch
Next
to continue. If the color has changed to yellow touch
Yellow
and refer to the “Troubleshooting“ section of this guide (Chapter 8).
Figure 7.
Chemical Indicator Color Change
9
Note:
Chemical indicator confirmation for a positive control BI can be changed in
the Utilities Menu under Control Alerts.