In-Progress Summary
Interpretation of Results
This section explains the different types of results, what each result means, and
what should be done after receiving each result.
Chemical Indicator Result
Read the chemical indicator (CI) result by comparing the CI color against the CI of
an unprocessed BI. The CI shifts from red/pink (unprocessed) to yellow (processed) after
exposure to the hydrogen peroxide sterilization cycle. A processed CI may be yellow or
yellow with some red/orange/brown dots.
Figure 17.
Chemical Indicator Color Change
Yellow: color
change
Red / Pink: no
color change