Load Preparation
3
STERRAD
®
100NX
®
User’s Guide
23
Thermosetting Elastomers
Silicone
Polyurethane
PVC
Glass
Glass
Metal
Aluminum
Brass
Gold
Stainless steel
Titanium
Items Not To Be Processed
Single use items for which the manufacturer does not recommend
resterilization.
Liquids and powders.
Items or materials that absorb liquids.
Items made of materials that contain cellulose, such as cotton, paper or
cardboard, linens, huck towels, gauze sponges, or any item containing
wood pulp
Paper instrument count sheets or lot stickers.
Items with hinged/mated nylon surfaces.
Instruments and devices that cannot withstand a vacuum and are labeled
for gravity steam sterilization methods only.
Items whose design permits the surfaces to collapse onto each other
unless some method is used to keep the surfaces separated.
Devices with dead-end lumens.
DRAFT
Содержание Sterrad 100NX
Страница 1: ...STERRAD 100NX Sterilization System User s Guide Ref 99994 M 99994_03 May 2014 D R A F T...
Страница 21: ...Load Preparation 3 STERRAD 100NX User s Guide 19 D R A F T...
Страница 22: ...3Load Preparation 20 STERRAD 100NX User s Guide D R A F T...
Страница 23: ...Load Preparation 3 STERRAD 100NX User s Guide 21 D R A F T...
Страница 49: ...Operation 4 STERRAD 100NX User s Guide 47 Cycle Completion Flowchart D R A F T...
Страница 50: ...4Operation 48 STERRAD 100NX User s Guide STERRAD CYCLESURE 24 Biological Indicator Flowchart D R A F T...
Страница 76: ...7Maintenance 74 STERRAD 100NX User s Guide D R A F T...
Страница 80: ...8Reports and Files 78 STERRAD 100NX User s Guide D R A F T...
Страница 104: ...9Access Levels and Supervisor Tasks 102 STERRAD 100NX User s Guide D R A F T...
Страница 108: ...A Specifications 106 STERRAD 100NX User s Guide D R A F T...