Load Preparation
3
STERRAD
®
100NX
®
User’s Guide
25
Guidelines for Preparing Items to
Be Sterilized
Note: All items must be cleaned, rinsed, and thoroughly
dried before being placed in the STERRAD
®
100NX
®
Sterilizer. Loads containing moisture may cause
cycle cancellations.
Cleaning, Rinsing, and Drying
Cleaning and sterilization are two separate processes. Proper cleaning of
instruments and devices is a critical and necessary step prior to sterilization.
All items including accessories must be thoroughly cleaned, rinsed, and
dried before loading into the sterilizer.
Carefully inspect all instruments, devices, and accessories for cleanliness
and dryness prior to packaging. If visible soil is present, the item must be
re-cleaned and dried prior to sterilization. If moisture is present, dry the
item thoroughly prior to sterilization.
Carefully inspect all instruments, devices, and accessories for flaws or
damage prior to packaging. Items with flaws or damage should be
replaced or repaired before using.
Note: Periodic careful inspection of items after repeated
exposure to disinfectant/cleaner/sterilant is necessary,
due to the potential damaging effects of the chemical
agents.
Cleaning is necessary to remove organic and inorganic soil and debris from
equipment. This process also removes many microorganisms from the surface
of the items. Sterilization then inactivates all remaining spores and live
microorganisms.
Clean
your devices according to the medical device manufacturers'
instructions. You must remove all blood, tissue, and soil from items using
appropriate detergents, cleansers and/or methods.
Rinse
items thoroughly to remove detergent or cleanser residue. Use
treated water that is of a quality that ensures hard water stains do not
occur. Failure to remove all organic materials or detergents may result in
the formation of light-colored residue on the devices. If residue is visible,
you should clean, rinse, dry, and resterilize the device prior to use.
DRAFT
Содержание Sterrad 100NX
Страница 1: ...STERRAD 100NX Sterilization System User s Guide Ref 99994 M 99994_03 May 2014 D R A F T...
Страница 21: ...Load Preparation 3 STERRAD 100NX User s Guide 19 D R A F T...
Страница 22: ...3Load Preparation 20 STERRAD 100NX User s Guide D R A F T...
Страница 23: ...Load Preparation 3 STERRAD 100NX User s Guide 21 D R A F T...
Страница 49: ...Operation 4 STERRAD 100NX User s Guide 47 Cycle Completion Flowchart D R A F T...
Страница 50: ...4Operation 48 STERRAD 100NX User s Guide STERRAD CYCLESURE 24 Biological Indicator Flowchart D R A F T...
Страница 76: ...7Maintenance 74 STERRAD 100NX User s Guide D R A F T...
Страница 80: ...8Reports and Files 78 STERRAD 100NX User s Guide D R A F T...
Страница 104: ...9Access Levels and Supervisor Tasks 102 STERRAD 100NX User s Guide D R A F T...
Страница 108: ...A Specifications 106 STERRAD 100NX User s Guide D R A F T...