Load Preparation
3
STERRAD
®
100NX
®
User’s Guide
17
EXPRESS Cycle Processing
The following types of medical devices can be sterilized in the
EXPRESS
Cycle:
General medical devices requiring surface sterilization, or sterilization of
mated titanium and stainless steel surfaces.
Rigid or semi-rigid endoscopes without lumens; for example da Vinci
®
endoscopes.
FLEX Cycle Processing
Medical devices, including most flexible endoscopes, with the following
materials and dimensions can be processed in the STERRAD
®
100NX
®
Sterilizer
FLEX cycle:
Single channel polyethylene and Teflon
®
(polytetrafluoroethylene)
flexible endoscope with an inside diameter of 1 mm or larger and length
of 850 mm or shorter.
**
Flexible endoscopes without lumens.
Note: Do Not process more than 2 flexible endoscope per
load.
Check the medical device manufacturer’s instructions before loading any
item into the STERRAD
®
100NX
®
Sterilizer.
**
One or two flexible endoscopes can be processed per sterilization cycle. No additional load.
DRAFT
Содержание Sterrad 100NX
Страница 1: ...STERRAD 100NX Sterilization System User s Guide Ref 99994 M 99994_03 May 2014 D R A F T...
Страница 21: ...Load Preparation 3 STERRAD 100NX User s Guide 19 D R A F T...
Страница 22: ...3Load Preparation 20 STERRAD 100NX User s Guide D R A F T...
Страница 23: ...Load Preparation 3 STERRAD 100NX User s Guide 21 D R A F T...
Страница 49: ...Operation 4 STERRAD 100NX User s Guide 47 Cycle Completion Flowchart D R A F T...
Страница 50: ...4Operation 48 STERRAD 100NX User s Guide STERRAD CYCLESURE 24 Biological Indicator Flowchart D R A F T...
Страница 76: ...7Maintenance 74 STERRAD 100NX User s Guide D R A F T...
Страница 80: ...8Reports and Files 78 STERRAD 100NX User s Guide D R A F T...
Страница 104: ...9Access Levels and Supervisor Tasks 102 STERRAD 100NX User s Guide D R A F T...
Страница 108: ...A Specifications 106 STERRAD 100NX User s Guide D R A F T...