Safety Information
2
STERRAD
®
100NX
®
User’s Guide
9
WARNING! RISK OF BREATHING DIFFICULTIES
On rare occasions, the outlet filter on the vacuum pump can prematurely fail. If this
occurs, you may see mist or what some users have described as “haze” or “smoke” in
the room where the sterilizer is operating. The chemical composition of the mist is
primarily airborne mineral oil with trace amounts of other compounds. Oil mist
exposure may, theoretically, pose an increased risk to people with certain respiratory
conditions, such as asthma, and they should take special precautions not to be exposed
to the mist. If you observe these conditions, personnel should leave the room as a
precaution and discontinue use of the STERRAD
®
System until the system is
repaired. Personnel should avoid working in the room until the mist has cleared.
Please note that all STERRAD
®
Sterilizers should be used and installed in a well-
ventilated environment (a minimum of 10 air exchanges per hour).
Personal Protective Equipment
WARNING! HYDROGEN PEROXIDE MAY BE PRESENT
Wear chemical resistant latex, PVC (vinyl), or nitrile gloves whenever handling a load
after a cycle cancellation. Hydrogen peroxide liquid may be present on the load or in
the chamber.
Device Safety
WARNING! RISK OF INJURY OR DAMAGE TO STERILIZER
The STERRAD
®
100NX
®
Sterilizer should not be used stacked with other
equipment.
CAUTION: RISK OF DAMAGE TO LOAD
Metal objects must not come into contact with the chamber walls, the door, or the
electrode. Contact with the walls, door, or electrode could damage the sterilizer or
the metal objects.
DRAFT
Содержание Sterrad 100NX
Страница 1: ...STERRAD 100NX Sterilization System User s Guide Ref 99994 M 99994_03 May 2014 D R A F T...
Страница 21: ...Load Preparation 3 STERRAD 100NX User s Guide 19 D R A F T...
Страница 22: ...3Load Preparation 20 STERRAD 100NX User s Guide D R A F T...
Страница 23: ...Load Preparation 3 STERRAD 100NX User s Guide 21 D R A F T...
Страница 49: ...Operation 4 STERRAD 100NX User s Guide 47 Cycle Completion Flowchart D R A F T...
Страница 50: ...4Operation 48 STERRAD 100NX User s Guide STERRAD CYCLESURE 24 Biological Indicator Flowchart D R A F T...
Страница 76: ...7Maintenance 74 STERRAD 100NX User s Guide D R A F T...
Страница 80: ...8Reports and Files 78 STERRAD 100NX User s Guide D R A F T...
Страница 104: ...9Access Levels and Supervisor Tasks 102 STERRAD 100NX User s Guide D R A F T...
Страница 108: ...A Specifications 106 STERRAD 100NX User s Guide D R A F T...