Troubleshooting
52
STERRAD
®
100NX
®
User’s Guide
WARNING! HYDROGEN PEROXIDE MAY BE PRESENT.
Wear chemical resistant latex, PVC (vinyl), or nitrile gloves whenever handling
a load after a cycle cancellation. Hydrogen peroxide liquid may be present on
the load or in the chamber
.
H
2
O
2
Rate Outside the
Calibrated Range
Reboot the system. If the problem persists, call your ASP
Representative.
H
2
O
2
Sensor Fault
Reboot the system. If the problem persists, call your
ASP Representative.
High Plasma Power
The plasma power is out of specification. Remove the load and
run diagnostics.
Invalid Key Code. Please Enter A
Valid Code
Press the OK button and reenter a valid key code.
Invalid Load Items File
Press the Back button and reinsert the USB memory stick.
Less Number of Cells Available, Please
Dispose the Cassette
The system will automatically move the cassette to the
disposal box.
WARNING! HYDROGEN PEROXIDE MAY BE PRESENT.
Wear chemical resistant latex, PVC (vinyl), or nitrile gloves whenever handling
a load after a cycle cancellation. Hydrogen peroxide liquid may be present on
the load or in the chamber.
Load May Contain H
2
O
2
Residual peroxide may be present on the load or chamber walls
due to a cancelled cycle or system malfunction. Wearing
chemical resistant latex, PVC (vinyl), or nitrile gloves,
repackage the load and restart the cycle. If the problem persists,
call your ASP Representative.
WARNING! HYDROGEN PEROXIDE MAY BE PRESENT.
Wear chemical resistant latex, PVC (vinyl), or nitrile gloves whenever handling
a load after a cycle cancellation. Hydrogen peroxide liquid may be present on
the load or in the chamber.
Low Plasma Power
The plasma power is out of specification. Remove the load and
run diagnostics.
Memory Card Full
Initiate backup to the memory card or press OK and the system
will delete the oldest file.
No File Found
Press OK. Reinsert the USB memory stick.
Other Door Is Open
Close the door on the other side of the sterilizer.
Please Close Door
Touch the close door display or the door foot pad to close
the door.
If this message appears. . .
Do this . . .
DRAFT
Содержание Sterrad 100NX
Страница 1: ...STERRAD 100NX Sterilization System User s Guide Ref 99994 M 99994_03 May 2014 D R A F T...
Страница 21: ...Load Preparation 3 STERRAD 100NX User s Guide 19 D R A F T...
Страница 22: ...3Load Preparation 20 STERRAD 100NX User s Guide D R A F T...
Страница 23: ...Load Preparation 3 STERRAD 100NX User s Guide 21 D R A F T...
Страница 49: ...Operation 4 STERRAD 100NX User s Guide 47 Cycle Completion Flowchart D R A F T...
Страница 50: ...4Operation 48 STERRAD 100NX User s Guide STERRAD CYCLESURE 24 Biological Indicator Flowchart D R A F T...
Страница 76: ...7Maintenance 74 STERRAD 100NX User s Guide D R A F T...
Страница 80: ...8Reports and Files 78 STERRAD 100NX User s Guide D R A F T...
Страница 104: ...9Access Levels and Supervisor Tasks 102 STERRAD 100NX User s Guide D R A F T...
Страница 108: ...A Specifications 106 STERRAD 100NX User s Guide D R A F T...