Access Levels and Supervisor Tasks
9
STERRAD
®
100NX
®
User’s Guide
83
System Configuration
Use the System Configuration screen to set sterilizer features. Selections
on this screen allow you to set the volume of the alarm loudspeaker, the
language used in displays and reports, and several access, report, and
connection features. The sterilizer comes configured with factory-set defaults.
If you want to change the default settings, select your preferred settings.
Figure 41. System Configuration.
Access Control Option
User Login
requires that a user identification and password be entered before
the sterilizer can be loaded and run. This is the factory default setting.
No User Login
allows any person to operate the sterilizer.
IMS
Enabled
causes the system to capture data with an IMS system (optional).
Disabled
causes the system to not capture data with an IMS system
(optional).
Vacuum Units
Torr/mTorr
expresses vacuum measurements in torr and mtorr.
kPa/Pa
expresses vacuum measurements in kilopascals and Pascals.
This is the factory default setting.
DRAFT
Содержание Sterrad 100NX
Страница 1: ...STERRAD 100NX Sterilization System User s Guide Ref 99994 M 99994_03 May 2014 D R A F T...
Страница 21: ...Load Preparation 3 STERRAD 100NX User s Guide 19 D R A F T...
Страница 22: ...3Load Preparation 20 STERRAD 100NX User s Guide D R A F T...
Страница 23: ...Load Preparation 3 STERRAD 100NX User s Guide 21 D R A F T...
Страница 49: ...Operation 4 STERRAD 100NX User s Guide 47 Cycle Completion Flowchart D R A F T...
Страница 50: ...4Operation 48 STERRAD 100NX User s Guide STERRAD CYCLESURE 24 Biological Indicator Flowchart D R A F T...
Страница 76: ...7Maintenance 74 STERRAD 100NX User s Guide D R A F T...
Страница 80: ...8Reports and Files 78 STERRAD 100NX User s Guide D R A F T...
Страница 104: ...9Access Levels and Supervisor Tasks 102 STERRAD 100NX User s Guide D R A F T...
Страница 108: ...A Specifications 106 STERRAD 100NX User s Guide D R A F T...