4
Operation
44
STERRAD
®
100NX
®
User’s Guide
2. Touch
No
to continue with the cycle. Touch
Yes
to cancel the cycle. Once
the cycle cancellation sequence begins, the screen turns red and the
cancellation sequence cannot be interrupted. The cancellation sequence
may take up to ten minutes to complete.
Figure 15. Cycle Cancellation In Progress. Cancellation Has Been Confirmed.
Loads from canceled cycles should be rewrapped using new packaging
materials, STERRAD
®
Chemical Indicator Strips, and STERRAD
®
SEALSURE
®
Chemical Indicator Tape. If a biological indicator was used
in the canceled load, the previously used biological indicator must be
discarded and a new biological indicator must be placed in the chamber
before restarting the new cycle.
WARNING! HYDROGEN PEROXIDE MAY BE PRESENT.
If a cycle cancels and the items in the load appear wet, hydrogen peroxide may be
present. Wear chemical resistant latex, PVC (vinyl), or nitrile gloves while removing
the items from the chamber, and while wiping off the items with a damp cloth.
Discard contaminated cloth according to your facility’s procedures.
DRAFT
Содержание Sterrad 100NX
Страница 1: ...STERRAD 100NX Sterilization System User s Guide Ref 99994 M 99994_03 May 2014 D R A F T...
Страница 21: ...Load Preparation 3 STERRAD 100NX User s Guide 19 D R A F T...
Страница 22: ...3Load Preparation 20 STERRAD 100NX User s Guide D R A F T...
Страница 23: ...Load Preparation 3 STERRAD 100NX User s Guide 21 D R A F T...
Страница 49: ...Operation 4 STERRAD 100NX User s Guide 47 Cycle Completion Flowchart D R A F T...
Страница 50: ...4Operation 48 STERRAD 100NX User s Guide STERRAD CYCLESURE 24 Biological Indicator Flowchart D R A F T...
Страница 76: ...7Maintenance 74 STERRAD 100NX User s Guide D R A F T...
Страница 80: ...8Reports and Files 78 STERRAD 100NX User s Guide D R A F T...
Страница 104: ...9Access Levels and Supervisor Tasks 102 STERRAD 100NX User s Guide D R A F T...
Страница 108: ...A Specifications 106 STERRAD 100NX User s Guide D R A F T...