Operation
4
STERRAD
®
100NX
®
User’s Guide
37
2. When placing the load on the shelves, make certain that you do not
block the ultraviolet lamp beam in the front right (your right) side of
the chamber. Make sure the load is centered on the shelves and that the
shelves are centered in the chamber.
Figure 6. The Load Should NOT Touch the Electrode.
3. Do not allow any part of the load to touch the electrode, the back wall of
the chamber, or the inside of the door.
4. Leave at least 1 inch (25 mm) of free space between the load and the
electrode to allow hydrogen peroxide to diffuse around the load.
Figure 7. Do NOT Stack Trays.
DRAFT