Operation
4
STERRAD
®
100NX
®
User’s Guide
41
System Ready Screen
The System Ready screen displays a number of buttons allowing you to select
cycles based on the cycles that are loaded and enabled on your sterilizer.
Cycle Types
STANDARD
cycle sterilizes the load in about 47 minutes.
This cycle is
installed on all systems.
DUO
cycle is for sterilization of flexible endoscopes with their light cords
and cameras.It sterilizes the load in about 60 minutes.
FLEX
cycle is specifically designed for flexible endoscopes and sterilizes
the load in about 42 minutes.
EXPRESS
cycle is for surface sterilization of general medical devices,
rigid and semi-rigid endoscopes without lumens; and sterilization of
mated stainless steel and titanium devices. It sterilizations the load in
about 24 minutes.
Functions
Logout
is used when the current operator is finished using the sterilizer
and the option is enabled. When
Logout
is selected, you must re-login to
use the sterilizer.
Cycle History
displays the Select Cycle History screen. This screen
allows you to select a cycle history file and view or print it.
Utilities
are available only to operators with Supervisor-level access.
It displays the Additional Utilities Menu.
Door Open
opens the active door.
Door Close
closes the active door.
Inserting a Cassette
1. Take a new STERRAD
®
100NX
®
Cassette out of the shipping carton.
2. Look at the package carefully before opening it. The indicator strip
should be white.
If the indicator strip is red, or if you see droplets
of moisture, do not open the package
– it is possible that hydrogen
peroxide has leaked inside the package. Refer to the cassette
Instructions
for Use
for proper handling instructions.
3. If the indicator strip is white, open the cassette package.
4. Position the cassette so that the arrows are pointing towards the sterilizer.
DRAFT
Содержание Sterrad 100NX
Страница 1: ...STERRAD 100NX Sterilization System User s Guide Ref 99994 M 99994_03 May 2014 D R A F T...
Страница 21: ...Load Preparation 3 STERRAD 100NX User s Guide 19 D R A F T...
Страница 22: ...3Load Preparation 20 STERRAD 100NX User s Guide D R A F T...
Страница 23: ...Load Preparation 3 STERRAD 100NX User s Guide 21 D R A F T...
Страница 49: ...Operation 4 STERRAD 100NX User s Guide 47 Cycle Completion Flowchart D R A F T...
Страница 50: ...4Operation 48 STERRAD 100NX User s Guide STERRAD CYCLESURE 24 Biological Indicator Flowchart D R A F T...
Страница 76: ...7Maintenance 74 STERRAD 100NX User s Guide D R A F T...
Страница 80: ...8Reports and Files 78 STERRAD 100NX User s Guide D R A F T...
Страница 104: ...9Access Levels and Supervisor Tasks 102 STERRAD 100NX User s Guide D R A F T...
Страница 108: ...A Specifications 106 STERRAD 100NX User s Guide D R A F T...