4
Operation
32
STERRAD
®
100NX
®
User’s Guide
For the STERRAD
®
100NX
®
DUO cycle in the United States, the
STERRAD
®
CYCLESURE 24 BI should only be used in a test pack
configuration. Instructions for use are provided in the test pack.
Login
Note: If your sterilizer has been configured not to require
operator login, the login screen will not appear. Skip to
the subsection titled Entering Load Information.
When you touch the “Touch Screen to Start” screen, the sterilizer displays the
Operator Login screen.
Figure 1. Operator Login Screen.
1. Touch the
Operator
field. The cursor appears in the field.
Note: Operator and Password fields are case-sensitive.
2. Use the on-screen keyboard to type your assigned operator identification.
3. Touch the
ENTER
key. The cursor jumps to the
Password
field.
4. Type your password. The screen displays a series of “*” characters in
place of the characters you type. This is done to keep others from reading
your password.
5. When you have finished entering your password, touch the
ENTER
key.
DRAFT
Содержание Sterrad 100NX
Страница 1: ...STERRAD 100NX Sterilization System User s Guide Ref 99994 M 99994_03 May 2014 D R A F T...
Страница 21: ...Load Preparation 3 STERRAD 100NX User s Guide 19 D R A F T...
Страница 22: ...3Load Preparation 20 STERRAD 100NX User s Guide D R A F T...
Страница 23: ...Load Preparation 3 STERRAD 100NX User s Guide 21 D R A F T...
Страница 49: ...Operation 4 STERRAD 100NX User s Guide 47 Cycle Completion Flowchart D R A F T...
Страница 50: ...4Operation 48 STERRAD 100NX User s Guide STERRAD CYCLESURE 24 Biological Indicator Flowchart D R A F T...
Страница 76: ...7Maintenance 74 STERRAD 100NX User s Guide D R A F T...
Страница 80: ...8Reports and Files 78 STERRAD 100NX User s Guide D R A F T...
Страница 104: ...9Access Levels and Supervisor Tasks 102 STERRAD 100NX User s Guide D R A F T...
Страница 108: ...A Specifications 106 STERRAD 100NX User s Guide D R A F T...