OVER
Steps
Do This
6.
Prepare the insertion site.
a. Wash your hands thoroughly with soap and water.
b. Clean the selected insertion site area with soap
and water.
c. Rinse the area thoroughly and pat dry.
d. Wipe the area with an alcohol prep pad.
Note 1: The insertion area MUST be clean and dry.
Otherwise:
• An infection could occur.
• The sensor support mount may not stick to the site.
Note 2: Do NOT place a bandage on the insertion
site before inserting the sensor. The sensor may not
penetrate the skin and the system will not work with a
bandage under the sensor.
7.
Remove the sensor delivery
unit from package.
a. Remove the sensor delivery unit from its sterile package
by peeling off the foil on the back of the package.
b. Save the package.
Note: You will need the sensor code number on the back
of the package, later.
Important:
We recommend that you save the sensor kit
carton or the sensor sterile package until the last sensor
from your sensor kit has been used and discarded. Saving
the package will help you to have sensor lot information,
expiration dates and sensor code etc. accessible when you
need them.
8.
Remove the protective liner from
the sensor support mount.
a. Bend the two-piece protective liner slightly so you can
see the seam between the two pieces.
b. Peel away the smaller section.
c. Peel away the larger section.
Caution:
Do
NOT
use the sensor delivery unit if the
sterile package is open or damaged.
Section 5
39
Insert or Remove Your Sensor
Содержание FREESTYLE NAVIGATOR
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 9: ...1 Getting Acquainted ...
Страница 23: ...2 Install or Replace Transmitter and Receiver Batteries ...
Страница 29: ...3 Set the Time and Date ...
Страница 34: ...Customer Care 1 866 597 5520 ...
Страница 35: ...4 Perform a Control Solution Test ...
Страница 42: ...Customer Care 1 866 597 5520 ...
Страница 43: ...5 Insert or Remove Your Sensor ...
Страница 54: ...Customer Care 1 866 597 5520 ...
Страница 55: ...6 Attach Your Transmitter ...
Страница 61: ...7 Calibrate Your System Test Blood Glucose Manually ...
Страница 74: ...Customer Care 1 866 597 5520 ...
Страница 75: ...8 Set Review or Change the Alarm Settings ...
Страница 89: ...9 Daily Use ...
Страница 96: ...Customer Care 1 866 597 5520 ...
Страница 97: ...10 Respond to Alarms Errors and Problems ...
Страница 119: ...11 Add Events ...
Страница 125: ...12 Review Reports and Edit Events ...
Страница 135: ...13 Clean Your System ...
Страница 136: ...Section 13 Key Terms n Receiver n Transmitter n Contact Points Customer Care 1 866 597 5520 ...
Страница 139: ...14 Link and Unlink Your Transmitter and Receiver ...
Страница 140: ...Section 14 Key Terms n BG Mode n CM Mode n Link Unlink n Receiver n Transmitter Customer Care 1 866 597 5520 ...
Страница 144: ...Customer Care 1 866 597 5520 ...
Страница 145: ...15 Reset User Settings ...
Страница 146: ...Customer Care 1 866 597 5520 ...
Страница 149: ...16 Change Country Settings ...
Страница 155: ...17 Appendix A Site Maintenance ...
Страница 158: ...Customer Care 1 866 597 5520 ...
Страница 159: ...18 Appendix B Specifications ...
Страница 173: ...19 Appendix C FCC Compliance Information ...
Страница 174: ...Customer Care 1 866 597 5520 164 Section 18 Appendix B Specifications ...
Страница 182: ...Customer Care 1 866 597 5520 ...
Страница 183: ...20 Glossary ...
Страница 184: ...Customer Care 1 866 597 5520 174 Section 19 Appendix C Classification Compliance Declaration ...
Страница 191: ...21 Index ...
Страница 192: ...Customer Care 1 866 597 5520 ...
Страница 195: ......