Setup is used to specify the triggering device and method for logging data.
B.3.1
Data Organization
The extended memory option organizes data recordings as a set of one or
more files – up to 20. Each file consists of one header record and a set
of data records. The first record, with the header information, resides in
standard memory, and subsequent records reside in extended memory. As
new records are logged, they will be sent into this file unless a new Data
Type is specified, or a new Log is begun (in the Auto-Log Setup).
When the Model 931A creates a header record, it contains two
types of information: measurement data and instrument configuration
data. Measurement data consists of everything described in the following
paragraph under Normal or Custom Data types. Whereas configuration
data contains everything necessary to reconfigure the instrument exactly
as it was at the time of recording.
Two categories of data may be recorded to extended memory: Normal
and Custom. Normal data consists of a complete set of data parameters,
whereas the custom data function allows the user to select the individual
parameters to be recorded, changing the record structure. The custom
data feature can be used to maximize the number of data records stored
in extended memory.
B.3.1.1
Normal
Each record stored as normal data contains enough information to com-
pletely restore the Model 931A to its state when the record was logged.
The type of data stored depends on the active instrument display at the
time the record was logged. There are four types of normal data:
Standard data
includes all of the Narrow-Band and Wide-Band
parameters for Channels 1 and 2, along with Timer and Transducer
results. Press the LOG DATA key when viewing any of the stan-
dard data formats (i.e.
VOLTS/AMPS, PHASE/FREQUENCY,
POWER/VARS, VA/PF). All logged data is tagged with the time
and date. This type of record occupies the least memory space. One
record of Standard data occupies 153 bytes of memory.
Harmonic data
incorporates all of the features of standard data but
includes harmonic information from Channel 1 and 2 signals. Press
the LOG DATA key anytime while displaying harmonics to record
130
Содержание 931A
Страница 6: ...vi ...
Страница 12: ...xii ...
Страница 14: ...xiv ...
Страница 16: ...xvi ...
Страница 52: ...SOURCE LOAD B A C N Figure 3 12 3 Phase 4 Wire 3 Element Connections 36 ...
Страница 90: ...Iout I1 V1 V2 V3 I2 I3 Figure 3 17 3 Phase 4 Wire 3 Element Transducer Test 74 ...
Страница 114: ...98 ...
Страница 120: ...104 ...
Страница 142: ...126 ...