A.3.17
Harmonic Phase Angle, Return (HPH)
Command:
harm#,ch#HPH
Returns individual harmonic phase angle for the selected channel. Set
the unit to measure harmonics, on the selected channel using the HAR
command, prior to issuing this command.
harm#
To select
2–50
Harmonic number (2 through 50)
1
THD (Total harmonic distortion)
ch#
1
Channel 1
2
Channel 2
Response: Numeric ASCII string.
A.3.18
Power Down (KIL)
Command:
KIL
Turns unit OFF.
A.3.19
Log Data (LOG)
Command:
LOG
Stores the current measurement data (using current date, time and
instrument settings) in the next available location of the non-volatile data
memory. Stores all data with current data type (standard data, standard
with harmonic data, or standard with harmonic and waveform data).
Recall data using the review logged data (#RLD) command.
A.3.20
Last Setup (LSU)
Command:
LSU
Restores the instrument to the state it was in immediately prior to the
last change. Use this command to undo an unwanted change, or to toggle
between two different setups. When this command is sent, the current
state is also stored as the new last setup.