s/r#
0
Set Calibration Constant
1
Return Calibration Constant
Return format:
gain < cr >< lf >
phase/offset
< cr >< lf >
. Numbers
are returned in fixed-point format with 5 digits to the right of the decimal
point.
A.3.4
Ratio CH1/CH2 (CCR)
Command:
CCR
Displays the measured results (on the Model 931A) as the magnitude and
phase angle of the ratio of Channel 1 divided by Channel 2. Depending on
the input signals selected, the display units for magnitude will be ohms,
mhos, or p/u.
A.3.5
Channel Input Select (CHS)
Command:
in#,ch#CHS
Specifies the input signal for the selected channel as follows:
AC
AC
DC
in#
Config.
in#
Config.
in#
Config.
0
Vab
11
Vcs
15
Vab
1
Vbc
12
Ia
16
Vbc
2
Vca
13
Ib
17
Vca
3
Vac
14
Ic
18
Vac
4
Vcb
27
Vb(Ch1 only)
19
Vcb
5
Vba
28
In (Ch2 only)
20
Vba
6
Van
21
Van
7
Vbn
22
Vbn
8
Vcn
23
Vcn
9
Vas
24
Vas
10
Vbs
25
Vbs
26
Vcs
Ch#
To select
1
Channel 1
2
Channel 2
108