3.8.2
Connecting the Inputs
If needed, use the main measurement inputs on the Model 931A as the
input for transducer tests at any value up to their rated maximum (see
Specifications in Section 4).
Connect the voltage and current measurement inputs according to the
type of transducer being tested. The operating software defines which
combination of inputs to use for testing of each type of transducer (except
1-Phase 2-Wire). These definitions are described in the Transducer Mode
Selection menu, shown in Section 3.8.1.1.
Figures 3.14 through 3.17 illustrate various examples of connections for
transducer testing.
Use any of the voltage and current inputs to measure single-phase, two-
wire transducers since they do not require sequential measurements. Use
channel 1 for voltage and channel 2 for current, as channel 1 is used as a
reference channel, as the voltage signal is typically the most stable signal.