3-4 Commands
Command Directory
3-14
PN: 10370-10374 Rev. F
MG369xC GPIB PM
ATT(xx)
Description:
Sets the step attenuator value to xx (times 10 dB).
Permits setting the step attenuator value to xx (times 10 dB) in the ALC step attenuator
decoupling mode.
Functional
Category:
POWER LEVELING
Value:
xx: unsigned integer between 00 (0 dB attenuation) and 11 (110 dB attenuation)
Example:
AT1; ATT08
Decouples the step attenuator from the ALC system and sets the step attenuator value to
80 dB.
Related
Commands:
AT0
,
AT1
Front Panel Key:
Level
|
ALC Mode
|
Attenuate >
|
Decouple
|
Increase Attenuation
or
Decrease
Attenuation
AUT
Description:
Selects automatic sweep triggering.
Selects automatic sweep triggering for frequency, power level, or list sweeps. When
automatic sweep triggering is selected, the sweep continually sweeps from its start
frequency, power level, or list index to its stop frequency, power level, or list index with
optimal retrace time.
Functional
Category:
SWEEP
Example:
SSP; SF1; AUT
Sets step sweep F1-F2 with auto triggering.
Related
Commands:
EXT
,
TEX
Front Panel Key:
Frequency
| Step Sweep
|
More >
|
Trigger >
|
Auto
Frequency
| List | Sweep > | Trigger > | Auto
Level
|
Level Sweep | Trigger > | Auto
BPN
Description:
Selects -3.3V for retrace and bandswitch blanking outputs.
Selects a -3.3V level for the rear panel retrace and bandswitch blanking outputs.
The retrace blanking output signal is available at pin six of the
AUX I/O
connector.
The bandswitch blanking output signal is available at pin 20 of the
AUX I/O
connector.
Functional
Category:
CONFIGURATION
Front Panel Key:
System
|
Config | Rear Panel > | Bl/-