Programming with GPIB Commands
2-9 Modulation Commands
MG369xC GPIB PM
PN: 10370-10374 Rev. F
2-27
IP
Turns on internal pulse modulation.
P0
Turns off the internal or external pulse modulation function (This is the default mode).
PC1
Selects the 10 MHz internal pulse clock rate.
PC4
Selects the 40 MHz internal pulse clock rate (Option 24); Selects the 100 MHz internal pulse clock
rate (Option 26).
PH0
Turns off the internal or external
M function.
PH1
Turns on the external
M function in Narrow mode.
PH2
Turns on the external
M function in Wide mode.
PH7
Turns on the internal
M function in Narrow mode.
PH8
Turns on the internal
M function in Wide mode.
PHN
Turns on the external
M function in Narrow mode (Same as PH1).
PHV(x)
Selects the internal
M waveform, where x = 1-sine wave, 2-square wave, 3-positive ramp,
4-negative ramp, 5-Gaussian noise, 6-uniform noise, 7-triangle wave.
PHW
Turns on the external
M function in Wide mode (Same as PH2).
PMD(x)
Selects the internal pulse mode, where x = 1-single, 2-doublet, 3-triplet, 4-quadruplet.
PTF
Selects internal pulse triggering on the falling edge of an external input. Active only when pulse
trigger is gated, triggered, or triggered with delay.
PTG(x)
Selects the internal pulse trigger, where x = 1-free run, 2-gated, 3-delayed, 4-triggered, 5-triggered
with delay, 6-composite.
PTR
Selects internal pulse triggering on the rising edge of an external input. Active only when pulse
trigger is gated, triggered, or triggered with delay.
SC0
Turns off the SCAN modulation function.
SC1
Turns on the SCAN modulation function.
SD0
Turns off the internal pulse stepped delay mode.
SD1
Turns on the internal pulse stepped delay mode.
SQP
Turns on internal 1 kHz square wave pulse modulation.
SW0
Turns off the internal or external pulse modulation function (Same as P0).
SW1
Turns on internal 400 Hz square wave pulse modulation.
SW2
Turns on internal 1 kHz square wave pulse modulation (Same as SQP).
SW3
Turns on internal 7.8125 kHz square wave pulse modulation.
SW4
Turns on internal 27.8 kHz square wave pulse modulation.
XP
Turns on the external pulse modulation function. Disables the internal pulse modulation function,
if previously programmed.
Table 2-9.
Modulation Command List (2 of 2)
Mnemonic
Code
Function