Command Directory
3-4 Commands
MG369xC GPIB PM
PN: 10370-10374 Rev. F
3-13
ASD <value> <unit>
Description:
Opens the external AM dB/V sensitivity parameter.
Functional
Category:
PARAMETER ENTRY
Value:
0 dB/V to 25 dB/V
Units:
DV
Example:
AM2; ASD 3 DV
Turns on external log AM with a sensitivity of 3 dB/Volt.
Output Command:
OAB
Related
Commands:
AM2
Front Panel Key:
Modulation
|
AM
(external) |
Edit Sensitivity
AT0
Description:
Deselects ALC step attenuator decoupling.
Deselects the ALC step attenuator decoupling function. This returns the MG369xC to its
normal (coupled) leveling mode. In this mode, when the desired power level is set, the
correct combination of ALC level and step attenuator setting is determined by the
instrument firmware.
Functional
Category:
POWER LEVELING
Related
Commands:
AT1
,
ATT(xx)
Front Panel Key:
Level
|
ALC Mode
|
Attenuate >
|
Decouple
AT1
Description:
Selects ALC step attenuator decoupling.
This function decouples the step attenuator (if equipped) from the ALC system.
This permits separate setting the ALC level and step attenuator.
Functional
Category:
POWER LEVELING
Related
Commands:
AT0
,
ATT(xx)
Front Panel Key:
Level
| ALC Mode | Attenuate > | Decouple