3-4 Commands
Command Directory
3-22
PN: 10370-10374 Rev. F
MG369xC GPIB PM
CM8 or CM8 <value> <unit>
Description:
Sets CW mode at M8 and opens the M8 parameter.
Places the MG369xC in CW mode at the preset (or previously set) M8 frequency and
causes M8 to be output. Also permits setting the M8 parameter. The frequency setting is
determined by the frequency range of the particular MG369xC model.
Functional
Category:
CW FREQUENCY
Value:
Frequency range of the MG369xC model
Units:
GH, MH, KH, HZ
Related Output
Command:
OM8
Commands:
M8
,
ACW
Front Panel Key:
Frequency
| CW
|
Frequency Control >
|
Frequency List. . .
| highlight
M8
, then press
Edit Selection
or
Output Frequency
CM9 or CM9 <value> <unit>
Description:
Sets CW mode at M9 and opens the M9 parameter.
Places the MG369xC in CW mode at the preset (or previously set) M9 frequency and
causes M9 to be output. Also permits setting the M9 parameter. The frequency setting is
determined by the frequency range of the particular MG369xC model.
Functional
Category:
CW FREQUENCY
Value:
Frequency range of the MG369xC model
Units:
GH, MH, KH, HZ
Output Command:
OM9
Related
Commands:
M9
,
ACW
Front Panel Key:
Frequency
| CW
|
Frequency Control >
|
Frequency List. . .
| highlight
M9
, then press
Edit Selection
or
Output Frequency
CS0
Description:
Turns off the CW ramp.
Functional
Category:
MISCELLANEOUS
Related
Commands:
CS1
Front Panel Key:
Frequency
| CW | CW Ramp > | CW Ramp
Содержание MG369 C Series