Command Directory
3-4 Commands
MG369xC GPIB PM
PN: 10370-10374 Rev. F
3-11
AMO
Description:
Turns off the internal AM function generator if it is on.
Functional
Category:
MODULATION
Related
Commands:
AM7
,
AM8
Front Panel Key:
N/A
AMR <value> <unit>
Description:
Opens the internal AM rate parameter.
Functional
Category:
PARAMETER ENTRY
Value:
0.1 Hz to 1 MHz for sine wave; 0.1 Hz to 100 kHz for square, triangle, and ramp
waveforms
Units:
MH, KH, HZ
Example:
AM7; AMR 10 KH
Turns on internal linear AM with a 10 kHz rate.
Output Command:
OAR
(in Hz)
Related
Commands:
AM7
,
AM8
Front Panel Key:
Modulation
|
AM
(internal) |
Edit Rate
AMS <value> <unit>
Description:
Opens the external AM %/V sensitivity parameter.
Functional
Category:
PARAMETER ENTRY
Value:
0 %/V to 100 %/V
Units:
PCV
Example:
AM1; AMS 30 PCV
Turns on external linear AM with a sensitivity of 30 percent/Volt.
Output Command:
OAS
Related
Commands:
AM1
Front Panel Key:
Modulation
|
AM
(external) |
Edit Sensitivity