3-4 Commands
Command Directory
3-82
PN: 10370-10374 Rev. F
MG369xC GPIB PM
OPS
Description:
Returns the power sweep number-of-steps to the controller.
Functional
Category:
OUTPUT
Related
Commands:
PNS
Front Panel Key:
N/A
OPT
Description:
Returns the name of the internal pulse trigger to the controller.
Possible returned values: “FREE RUN”, “GATED”, “DELAYED”, “TRIGGERED”, “TRIG
WITH DELAY”, “COMPOSITE”
Functional
Category:
OUTPUT
Front Panel Key:
N/A
OPW
Description:
Returns the internal pulse width1 value (in
s) to the controller (Same as OW1).
Functional
Category:
OUTPUT
Front Panel Key:
N/A
OSB
Description:
Returns the primary GPIB status byte to the controller.
Functional
Category:
OUTPUT
Front Panel Key:
N/A
OSD
Description:
Returns the step sweep dwell time (in ms) to the controller.
Functional
Category:
OUTPUT
Related
Commands:
SDT
Front Panel Key:
N/A
OSE
Description:
Returns the last GPIB syntax error to the controller.
Syntax errors are those that occur in the formulation of a program statement, such as
writing EXTTFS instead of EXTTRS.
Functional
Category:
OUTPUT
Front Panel Key:
N/A