Command Directory
3-4 Commands
MG369xC GPIB PM
PN: 10370-10374 Rev. F
3-101
SAM
Description:
Outputs both the current instrument setup and all stored instrument setups to the
controller.
The current instrument setup and all stored instrument setups are output to the
controller in a 41000-byte (approximately) binary data string. The controller stores the
data.
Functional
Category:
STORED SETUP
Related
Commands:
RCM
,
RCF
,
SAF
Front Panel Key:
N/A
SB0
Description:
Inhibits extended status byte 2 SRQ generation.
Inhibits an SRQ from being generated when the Extended Status Byte two bit is set.
This is the default mode.
Functional
Category:
STATUS
Front Panel Key:
N/A
SB1
Description:
Enables extended status byte 2 SRQ generation.
Enables an SRQ to be generated when primary status byte bit seven (extended status
byte two) is set and SQ1 has been programmed. The extended status byte two bit is set
whenever one (or more) of the unmasked status reporting functions in extended status
byte two is true.
Functional
Category:
STATUS
Example:
C code snippet to enable unlocked SRQ:
strcpy(buf, “csb sb1 mb2 sq1 el1");
buf[11] = 16; /*mb2 command mask value*/snd(source_addr, buf);
Related
Commands:
SQ1
,
MB2
Front Panel Key:
N/A
SC0
Description:
Turns on the SCAN modulation function.
If the SCAN Modulator was not installed, this command produces a syntax error.
Functional
Category:
MODULATION
Front Panel Key:
N/A
Содержание MG369 C Series
Страница 2: ......
Страница 6: ...Contents 4 PN 10370 10374 Rev F MG369xC GPIB PM ...
Страница 262: ...A 34 PN 10370 10374 Rev F MG369xC GPIB PM ...
Страница 264: ...Index 2 PN 10370 10374 Rev F MG369xC GPIB PM ...
Страница 265: ......