![Ankom TDF Скачать руководство пользователя страница 62](http://html.mh-extra.com/html/ankom/tdf/tdf_operators-manual_2944721062.webp)
Operator’s
Manual
pg.
62
Rev
A
4/30/12
Pump
Tube
Test
When you press the Pump Tube Test button on the Diagnostics screen, the following screen is displayed.
This diagnostic tests the integrity of the tubing from the pump to the output valves (IDF, SDF, and Waste). When you press the
START button, the following screen is displayed.
When you press one of the Continue buttons, the Pump Tube Test screen is once again displayed with the bottom part of the
screen showing the status of each tube. This diagnostic is automatically executed at the beginning of each new IDF/SDF or TDF
run.
Motor
Test
When you press the Motor Test button on the Diagnostics screen, the following screen is displayed.
←
Motor
Test
Set
Valves
GO
Motor
Direction
(FWD)
20
RPM
120
RPM
60
RPM
240
RPM
Motor
Speed
10.0
Amount
(ml)
All
lines
should
be
charged
before
test
is
run.
Verify
that
all
supply
containers
are
connected
and
contain
the
desired
fluid.
Cancel
Continue
(no
line
charge)
Continue
(line
charge)
←
Pump
Tube
Test
START
Status
Pressure
difference
(psi)
1
2
3
4
5
6
Содержание TDF
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 16: ...Operator s Manual pg 16 Rev A 4 30 12 This page intentionally left blank ...
Страница 30: ...Operator s Manual pg 30 Rev A 4 30 12 This page intentionally left blank ...
Страница 46: ...Operator s Manual pg 46 Rev A 4 30 12 This page intentionally left blank ...
Страница 56: ...Operator s Manual pg 56 Rev A 4 30 12 This page intentionally left blank ...
Страница 70: ...Operator s Manual pg 70 Rev A 4 30 12 This page intentionally left blank ...
Страница 72: ...Operator s Manual pg 72 Rev A 4 30 12 This page intentionally left blank ...