Operator’s
Manual
pg.
58
Rev
A
4/30/12
#
Name
Description
Result
Correction
E4 In-line
Heater
Over Temp
The device that heats the fluid before
delivery is reporting a temperature that is
higher than the specified range.
In-progress processes are
aborted.
Run diagnostics.
Replace sensor or
heater.
E5 Paddle
Heater
Under Temp
The Paddle Heater is reporting a temperature
lower than the specified range.
IDF process continues.
SDF process is
unaffected.
Failed heater is disabled.
Run diagnostics.
Replace sensor or
heater.
E6 In-line
Heater
Under Temp
(IDF Amylase
Phase)
The device that heats the fluid before
delivery is reporting a temperature that is
lower than the specified range for the
Amylase phase.
IDF process is aborted.
SDF process continues.
Run diagnostics.
Replace sensor or
heater.
E7 In-line
Heater
Under Temp
(IDF Water
Rinse)
The device that heats the fluid before
delivery is reporting a temperature that is
lower than the specified range for the IDF
Water Rinse phase.
IDF process is aborted.
SDF process has not yet
started.
Run diagnostics.
Replace sensor or
heater.
E8 In-line
Heater
Under Temp
(SDF EtOH95
delivery)
The device that heats the fluid before
delivery is reporting a temperature that is
lower than the specified range for the SDF
EtOH95 Delivery phase.
IDF process is aborted.
SDF process has not yet
started.
Run diagnostics.
Replace sensor or
heater.
E9 In-line
Heater
Cooling Failure
The device that heats the fluid before
delivery is reporting a temperature that is
higher than the specified range for the Water
Cooling phase.
In-progress processes are
aborted.
Run diagnostics.
Replace sensor or
heater.
E10 High
Line
Pressure
The pressure in one or more of the delivery
tubes is higher than the specified values.
In-progress processes are
aborted.
Check for plugged tube
or replace output valve.
E11 Low
Input
Pressure
The pressure at the input port is lower than
the specified range.
In-progress processes are
frozen in time.
The mixer is stopped.
The N
2
filter is disabled
until the fault is
corrected.
Check that N
2
supply is
connected and working.
E12 High
Input
Pressure
The pressure at the input port is higher than
the specified range.
Fault is reported.
All in-progress processes
continue.
Adjust high pressure
regulator gauge on the
instrument to 50psi.
E13 Tubing Failure
At least one of the pump tubes has broken
OR one of the output valves is not closing.
In-progress processes are
aborted.
Replace pump tubing
OR repair output valve.
Содержание TDF
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 16: ...Operator s Manual pg 16 Rev A 4 30 12 This page intentionally left blank ...
Страница 30: ...Operator s Manual pg 30 Rev A 4 30 12 This page intentionally left blank ...
Страница 46: ...Operator s Manual pg 46 Rev A 4 30 12 This page intentionally left blank ...
Страница 56: ...Operator s Manual pg 56 Rev A 4 30 12 This page intentionally left blank ...
Страница 70: ...Operator s Manual pg 70 Rev A 4 30 12 This page intentionally left blank ...
Страница 72: ...Operator s Manual pg 72 Rev A 4 30 12 This page intentionally left blank ...