3. Operating Procedure
(2) Turn ON the CONTROL keyswitch.
The laser is put to an operable status and the COOLER → ON !! screen appears.
How to see the displayed items
PUMP
Displays the ON/OFF status of the pump to circulate cooling water.
WATER TEMP
Indicates the secondary cooling water temperature. Upon completion of
SELF CHECK, a temperature measurement is started and the measured
value is indicated.
DEIONIZED
WATER RES
Indicates the resistivity of secondary cooling water.
NOT READY: Indicated when the resistivity is below 3.00 MΩ·cm.
READY:
Indicated when the resistivity is 3.00 MΩ·cm or more.
WATER
TEMPERATURE
Indicates the secondary cooling water temperature status.
LOW:
Displayed at 24ºC or less.
NORMAL:
Displayed at 25ºC to 40ºC.
After the power supply is turned on, the temperature of
cooling water is not stable for a while. Immediately after
the power supply is turned on, NORMAL is displayed when
the temperature rises up to 27ºC.
HIGH:
Indicated at 41ºC or more.
LASER POWER
MONITOR
Indicates the sensor status of the power monitor unit.
NOT READY: Indicated during warming-up.
READY:
Indicated at completion of warming-up.
When "DEIONIZED WATER RES" is (READY), "WATER TEMPERATURE" is
(NORMAL), and "LASER POWER MONITOR" is (READY), a high voltage is turned
ON and charging is started. Then, the HV → ON !! screen appears.
Содержание ML-2550A
Страница 1: ...AA06M0642E 37 OPERATION MANUAL YAG LASER WELDER ML 2550A 2551A...
Страница 19: ...Introduction Part...
Страница 20: ......
Страница 37: ...Installation and Preparation Part...
Страница 38: ......
Страница 46: ......
Страница 57: ...Operating Part...
Страница 58: ......
Страница 116: ......
Страница 144: ......
Страница 168: ......
Страница 173: ...Maintenance Part...
Страница 174: ......
Страница 203: ...Appendixes...
Страница 204: ......