4 Use
Note:
If it is started up without a patient category selection, the following settings are re-
tained:
•
Patient type (category, height and gender)
•
Ventilation mode (set-points, alarm thresholds)
4.8.11 Set-point selection
The
Set-points
function is used to select the set-points to be used according to application:
•
I:E, Ti/Ttot, Ti or Flow rate in VCV
•
I:E, Ti/Ttot or Ti in PCV
•
Tplat in % or seconds in VCV and SIMV
•
PS or PI in PSV and SIMV
•
RR min enabled or disabled in PSV
4.8.12 Display of the monitoring blocks
The
Monitoring
function allows the user to change the display order of the monitoring blocks.
The
Restoration
keys allow you to reset the order of the monitoring blocks.
4.9 Power supply sources
This ventilator is compatible with various electrical power sources:
•
AC power supply via the mains power supply device supplied with the machine;
•
Power supply from an external DC source;
•
Power supply from two internal direct-current sources (internal battery and extractable battery).
The electrical characteristics of each of these sources are described in Section Electrical specifi-
cations on page 112.
Note:
If the unit suddenly shuts down during ventilation (e.g. battery empty), when the unit is
reconnected to the external or auxiliary power and restarted, it automatically resumes ventila-
tion with the last parameters saved.
Note:
The alarm settings, history and observance are retained when the device is shut down.
Note:
When a complete power loss occurs with both AC mains or the auxiliary/internal bat-
teries, the machine history is preserved and can be displayed after the power supply has been
restored. This data is preserved due to a secondary internal battery back-up battery with a life
time of approximately ten years. Note that this secondary internal battery does not power any
ventilation modes. An alarm is triggered when the autonomy of the back-up battery becomes
too low. If this occurs, contact the technical department.
Note:
When the battery is completely exhausted, the machine is reset to the factory settings.
The alarm thresholds are reset and the alarm history is erased.
4.9.1 Managing the power supply
The supply board manages the energy source changeover automatically according to the following hi-
erarchy as follows:
73