9.2.1.2 Alarm sound level
Acoustic pressure level measured
Accuracy
High
Min : 62 dB (A)
Max : 97 dB (A)
Medium
Min : 60 dB (A)
Max : 87 dB (A)
low
Min : 46 dB (A)
Max : 69 dB (A)
± 3 dB (A)
Microphone position
1 meter, opposite the speaker (position where the acoustic pres-
sure is highest in the horizontal plane)
9.2.1.3 Electrical specifications
Main Power Supply
Input voltage
100 - 240 V AC (tolerance -10 % ; +10 %)
Frequency
50 to 60 Hz
Electrical power consumption
120 VA max. (0.12 kW)
Electrical class
II
Type
BF
Leak current
IEC 60601-1 compliant
Protection following a loss of power
Continuous audible alarm and patient vented to
atmosphere
Cable type
2-point mains cable having max. length of 2.50 m
Note: The use of cables other than those men-
tioned in this manual may lead to an increase in
emissions or a decrease in the immunity of the
device as specified in the following paragraph:
Electromagnetic Compatibility
11
Includes the high, medium, and low-priority conditions.
12
The values shown correspond to the device’s sound volume settings: Min corresponds to a set-
ting of 20% and Max corresponds to a setting of 100%.