5 Alarms and other messages
No.
Alarms
Priority
Alarm activation
Alarm activation delay
Actions
100 Excessive internal bat-
tery temperature!
HIGH
Mean temperature of
internal battery:
- Discharging: above
75°C or below -40°C
- Charging: above 70°C
or below -20°C
Immediate triggering
Use a different unit and
contact your technical
department
101 Excessive auxiliary bat-
tery temperature!
HIGH
Mean temperature of
auxiliary battery:
- Discharging: above
75°C or below -40°C
- Charging: above 70°C
or below -20°C
Immediate triggering
Replace the auxiliary
battery.
102 Excessive PCB supply
temperature!
HIGH
Power supply PCB tem-
perature above 95°C
Immediate triggering
Use a different unit and
contact your technical
department
103 Internal battery temper-
ature measurement er-
ror!!!
HIGH
Correct voltage at inter-
nal
battery terminals, but
zero current or tempera-
ture outside tolerance
Triggering within 10 s
Use a different unit and
contact your technical
department
104 Auxiliary battery temper-
ature measurement er-
ror!!!
HIGH
Correct voltage at auxil-
iary
battery terminals, but
zero current or tempera-
ture outside tolerance
Triggering within 10 s
Replace the auxiliary
battery.
105 Internal battery defec-
tive!
LOW
Actual capacity of the
internal battery too low
compared to its initial
capacity or battery older
than 2 years
Immediate triggering
Isolate the device. Ful-
ly discharge the inter-
nal battery, then fully
charge it.
If the alarm still occurs,
use a different unit and
contact your technical
department.
108 Inspiratory pause occlu-
sion in progress
None
(INFO cat-
egory)
Inspiratory pause
respi-
ratory function selected
Triggering at the start of
the expiratory phase
Not applicable.
109 Expiratory pause occlu-
sion in progress
None
(INFO cat-
egory)
Expiratory pause
respi-
ratory function selected
Triggering at the start of
the inspiratory phase
Not applicable.