5 Alarms and other messages
No.
Alarms
Priority
Alarm activation
Alarm activation delay
Actions
79
CO2 concentration out
of tolerance
MEDIUM
Measured CO2 > 15%
Immediate triggering
If CO2 actually < 15%
then calibrate the probe.
80
Calibrate the IRMA™
probe(CO2)
MEDIUM
Probe drift
Immediate triggering
Calibrate the probe.
81
IRMA™ (CO2) probe er-
ror
MEDIUM
Probe hardware or soft-
ware error
Immediate triggering
Disconnect and recon-
nect the probe. If the
problem persists, re-
place the probe.
82
CO2 measurement inop-
erative
MEDIUM
Probe not connected
Immediate triggering
Connect the IRMA
probe.
83
IRMA™ (CO2)probe: In-
ternal temperature out
of tolerance!
MEDIUM
The internal tempera-
ture of the probe has ex-
ceeded the maximum
measurement threshold
Immediate triggering
If the ambient tempera-
ture is normal, replace
the probe.
84
IRMA™ (CO2)probe: am-
bient pressure out of
tolerance!
MEDIUM
The atmospheric pres-
sure measured by the
probe has exceeded
the maximum measure-
ment threshold
Immediate triggering
If the atmospheric pres-
sure is normal, replace
the probe.
85
etCO2 high!!!
MEDIUM
etCO2 measurement
above the threshold set-
ting
Immediate triggering
Check the suitability of
alarm level settings with
the ventilation settings
in progress.
86
etCO2 low!!!
MEDIUM
etCO2 measurement be-
low the threshold set-
ting
Immediate triggering
Check the suitability of
alarm level settings with
the ventilation settings
in progress.
89
O2 mixer defect!!! Venti-
lation Effective to 21%
HIGH
Electronics failure
Inhibit if FiO2 set-point
= 21%
Immediate triggering
Set the FiO2 set-point =
21%
Use a different unit and
contact your technical
department
90
High oxygen supply
pressure!!!
HIGH
O2 supply
pressure
> 7 bar
Immediate triggering
Check pressure in sup-
ply network or cylinder
91
Loss of oxygen supply
pressure!!!
HIGH
O2 supply pressure is
too low
Triggering within 10 s
Check remaining O2 lev-
el in
cylinder or pressure in
supply network.
For operation on a low-
pressure source, ensure
that the high-pressure
O2 connection hose is
disconnected from the
machine.