3 Installation and commissioning
3.6 Assembly of the patient circuit and of the accessories
CAUTION:
Make sure that the use of accessories does not affect the safety and the expect-
ed performance of the device. Only use the accessories described below with the
Monnal T60
Advanced
.
CAUTION:
It is
necessary
to use a hydrophobic bacterial filter
(1)
on the ventilator’s inspira-
tory outlet
(2)
.
CAUTION:
An HME
(3)
(heat and moisture exchanger) filter can be used in addition to the hy-
drophobic bacteriological filter
(1)
. It must be placed on the Y-shaped part
(4)
.
3
4
CAUTION:
Air Liquide Medical Systems recommends using the patient circuits listed in
Chapter List of options and accessories on page 101. If not, the use of patient circuits con-
taining phthalates or bisphenol A poses risks for pregnant women, lactating women, and chil-
dren.
Monnal T60 Advanced
is compatible with double-branch, adult and pediatric circuits.
ry assembly cleaning protocol on page 141 then insert it into the housing until you here it click into
place.
Double-branch patient circuit
27
28
Connect the patient circuit to the ventilator and the humidifier (if used):
21