6 Maintenance
Accessories can be re-usable (autoclavable) or single-use (disposable).
Reusable elements must be regularly cleaned and disinfected to avoid any risk of infection.
This procedure, which is mandatory and extremely important, is the responsibility of the user.
6.1 Surface of the ventilator
Products
The following products may be used to regularly disinfect the external surfaces of
Monnal T60 Ad-
vanced
:
•
Mikrozid PAA wipes;
•
Mikrozid sensitive wipes;
•
Mikrozid universal wipes.
Note:
Any other equivalent disinfectant may be used as long as the product's manufacturer
demonstrates its compatibility with
Monnal T60 Advanced
.
CAUTION:
Follow the instructions of the product's manufacturer.
Surface cleaning and disinfection precautions
CAUTION:
Before the operation:
•
power off
Monnal T60 Advanced
;
•
insert the USB cover to stop liquid from entering the ventilator.
CAUTION:
During cleaning or disinfection: Do not let any liquid penetrate the inside of the
machine or come into contact with the connectors.
CAUTION:
After the operation, let the product dry completely before reusing
Monnal T60 Ad-
vanced
.
6.2 Bacteriological filter
Apply manufacturer recommendations concerning the frequency of replacement of the bacteriological
filter. Refer to the user instructions supplied with the filter.
CAUTION:
Reusing single-use accessories or consumables carries the risk of patient cross-
contamination.
CAUTION:
A clogged bacteriological filter can lead to an increase in inspiratory and expirato-
ry resistance.