4 Use
Note:
By default, FiO
2
monitoring and monitoring of the expiratory flow are enabled.
etCO
2
monitoring is enabled automatically when the CO
2
probe is connected to the ventilator.
When the CO
2
probe is connected or enabled, the CO
2
sensor calibrat. key
is used to calibrate the
probe.
CO
2
sensor calibration
The IRMA
™
probe must be calibrated whenever an offset is found in the measurements or when the
Calibrate CO
2
sensor
alarm is triggered.
The probe cannot be calibrated until approximately 10 seconds after it has been connected to the ven-
tilator.
Probe calibration must be performed with a new adapter on the sensor. This must be disconnected
from the Y-piece of the patient circuit and from the patient. Next, press the
CO
2
sensor calibrat
. key
to launch probe calibration. The green LED on the probe flashes for about five seconds during calibra-
tion.
Special care must be taken to avoid any breathing into the adapter during probe calibration. The pres-
ence of ambient air (21% O
2
and 0% CO
2
) in the probe is crucial for a successful calibration.
4.7.3 Patient monitoring
The
“Patient monitoring”
function is for accessing the patient’s data such as History (see History on
page 81), Trends (see section Trends on page 60) and the CPV report (see section CPV: Cardio-Pul-
monary Ventilation on page 45).
4.7.4 Configuration of the ventilator
Press the
Menu
key: the list of menu parameters is displayed.
Press the
Configuration
key.
Volume
To adjust the device’s audible alarms:
•
Press the
Volume
key,
•
Turn the control wheel to display the desired value,
•
Press the control wheel again to validate.