4 Use
Note:
For operation of the VT Target function, refer to PRVC mode.
CAUTION:
The set minimum frequency is a safety threshold below which the ventilator pro-
vides controlled ventilation based on the set patient frequency.
If the patient is disconnected, the frequency is no longer adapted until the patient is recon-
nected.
Safety features of the VT target function are applied when the function is activated.
Set-points
FiO2
inspired oxygen fraction (%)
PS
inspiratory assistance (cmH
2
O)
RR support
maintenance frequency at which the patient is ventilated when he or she no longer
breathes spontaneously (bpm)
PEEP
positive end of expiration pressure (cmH
2
O)
VT Target
target volume (mL)
I. Trig.
inspiratory trigger (L/min)
RR mini
minimum respiratory rate (bpm)
PI max
maximum inspiratory pressure (cmH
2
O)
Slope
inspiratory pressure slope (cmH
2
O/s)
E. Trig.
Expiratory trigger (% of inspiratory peak flow)
Note:
In PS-Pro mode, there is no apnea ventilation.
4.6.9 PSIMV (Synchronized Intermittent Mandatory Pressure Monitored Ventilation)
[PSIMV = Synchronized Intermittent Mandatory Pressure Monitored Ventilation]
Principle
PSIMV Mode combines mandatory assisted pressure-controlled ventilation and spontaneous patient
ventilation between the assisted controlled cycles.
Set-points
FiO2
inspired oxygen fraction
PI
inspiratory pressure (cmH
2
O)
RR SIMV
determines the frequency of the imposed cycles (bpm)
PEEP
positive end of expiration pressure (cmH
2
O)
Tins
inspiration time of intermittent controlled assisted cycles (s)
I.Trig
inspiratory trigger (L/min)
43