5 Alarms and other messages
No.
Alarms
Priority
Alarm activation
Alarm activation delay
Actions
59
Low respiratory rate!
LOW
Measured frequency be-
low the alarm threshold
Inhibition for 60 s on
ventilation start-up.
Inhibition during a
pause and 30 s after a
pause.
Triggering within 1 ven-
tilation cycle.
Check the suitability of
alarm level settings with
the ventilation settings
in progress.
60
VT or PI not reached!
Check settings
LOW
In volumetric mode, the
measured VTi is less
than 2/3 of the setting.
In pressure-controlled
mode, Ppeak is less
than 2/3 of the setting
Inhibition for 60 s on
ventilation start-up.
Inhibition during a
pause and 30 s after a
pause.
Triggering within 3 ven-
tilation cycles.
Check the ventilator set-
tings and patient circuit.
Check the air inlet and
the filter at the back of
the unit for obstruction.
Use a different unit and
contact your technical
department
62
High VTi!
LOW
Insufflated tidal volume
above the set threshold
Triggering within 1 ven-
tilation cycle.
Check the level of leak-
age
Check the suitability of
alarm level settings with
the ventilation settings
in progress.
In pressure-control ven-
tilation mode, when the
VTi high alarm thresh-
old is reached, the unit
switches to the expira-
tion phase.
63
High VTe!
LOW
Exhaled tidal volume
above the set threshold
Inhibition for 60 s on
ventilation start-up.
Triggering within 1 ven-
tilation cycle.
Check the suitability of
alarm level settings with
the ventilation settings
in progress.
64
High respiratory rate!
LOW
Measured
frequency above the
alarm threshold or au-
to-triggering of the in-
spiratory trigger
Inhibition for 60 s on
ventilation start-up.
Inhibition during a
pause and 30s after the
pause.
Triggering within 1 ven-
tilation cycle.
Check the suitability of
alarm level settings with
the ventilation settings
in progress.
Reduce the sensitivity of
the inspiratory trigger if
required.
65
Non Critical error! Venti-
lation Effective
LOW
Electronics failure
Immediate
triggering in stand-by
only
Use a different unit and
contact your technical
department
66
Non Critical error! Venti-
lation Effective
LOW
Electronics failure
Immediate
triggering in stand-by
only
Use a different unit and
contact your technical
department
89