Sensor programming
couple (e.g. connector ZA9400-FSx with Ntc). see 9.2.7, Man. 6.7.3).
´*T´ This defines a temperature sensor (Ntc, Pt100) as reference for tempera-
ture compensation. (see 9.2.5).
´*P´ This defines an atmospheric pressure sensor as reference for atmospheric
pressure compensation. (s. 9.2.6).
´#N´ This activates conversion to standard conditions for flow measuring.
(see 9.4.9) The remaining 8 characters can be used for the user’s own descrip-
tions.
´!´ at the end automatically indicates a specific linearization or calibration (see
12.3.11). This cannot be overwritten.
12.3.3 Averaging mode
The various averaging methods can be defined via the Averaging mode func-
tion; these are described in Section 11.4.2.
Function - No averaging :
Averaging mode : -----
Averaging over all active measuring point scans
CONT
Averaging over all measuring point scans in a cycle
CYCL
12.3.4 Locking the sensor programming
The functional parameters for each measuring point are protected by means of
the locking mode; this can be set to the desired locking level
(see Manual
6.3.12)
. Before programming you must lower the locking mode to an appropri-
ate level. If you see a dot in the display after the locking mode, this means that
this cannot be modified.
Locking level
Locked functions
0
None
1
measuring range + element flags + output mode
3
+ units
4
+ zero-point corr gain correction
5
+ base value + exponent
6
+ analog output, start and end
+ zero-point adjustment, temporary
7
+ limit values, maximum and minimum
Function
´Locking mode´:
Locking level : 5
In the menu
SENSOR PROGRAMMING
the functions are listed from top to bottom
in such a way that the locked functions cannot be selected.
ALMEMO
®
5690-2M
63