12. Programming using the programming menus
Active channels for min. cycle and available memory time
Meas. chan.: 24 active : 05
Enter cycle
(see 10.5, format hh:mm:ss.cc):
Cycle:
00:01:00.00
Minimal cycle
with 50
M/s
corresp. to active channels:
<MIN>
00:00:00.12
Cycle without saving and without sleep mode
Saving : - Standard: -
To select and activate
saving to memory
press
<ON>
Ø
Standard: -
To activate
sleep mode
(see 12.2.5) press
<ON>
Standard:
Ø
Available memory time from cycle and channel number
Memory time:
24d 13h
Measuring duration, after start; automatic stop after :
Meas. duration : 00:15:00
File name with memory connector (maximum 8 characters)
File name :
ALMEMO.001
Number e.g. room 12, measuring point 1 (see 12.2.3)
Number :
12-001 A
12.2.3 Numbering of measuring operations
To identify measurements or series of measurements these can be individually
numbered before starting. This number is output or saved when the next mea-
suring point scan starts. In this way individual measuring operations can be as-
signed to certain types of measurement or certain measuring points (see Man-
ual 6.7).
After selecting the function
Number
the 6-character number is entered as nor-
mal (see 10.5). You can use digits 0 to 9 and also the characters A, F, N, P,
and - or _ (space). The number is activated as soon as it has been entered; it
will then be followed by the letter ´
A
´ until the next cyclic or manual measuring
operation is saved.
Number function
(e.g. room 12, measuring point 1)
NUMBER:
12-001
A
To
zero-set
and deactivate the number press
<CLR>
To
activate
and
deactivate
the number press
<ON>
,
<OFF>
To
increment
and
activate
the number press
< +1 >
12.2.4 Starting and stopping measuring operations
A measuring operation can be started and
stopped not only by pressing the appropriate
keys but also using numerous other methods
provided in the START - STOP wizard menu.
Operation via the interface is described in the
Manual, Section 6.6. The function using a start
time and end time or measuring duration is de-
scribed in Section 12.1.4, limit value actions in
Section 12.4.3, and the relay and trigger vari-
ants in Section 12.6.2.
12.2.5 Scanning mode
For autonomous operation and / or for scanning by computer there are 4 scan-
58 ALMEMO
®
5690-2M
* ALMEMO 5690-2 *
¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯
MEASURING START - STOP
by pressing: START-STOP key
by interface comand: S2-X
by time ©
by limit values
by trigger cable
F ESC