12. Programming using the programming menus
To output sensor programming of all active
measuring points (command P15, see Man-
ual
6.2.3)
press
<PRINT>
12.3.1 Selecting the input channel
To view or edit a sensor's parameters you must first of all select the menu
SENSOR PROGRAMMING
and then set the required input channel by pressing
or
. Only sensors actually connected and channels actually acti-
vated can be processed. To activate new channels first press
<MALL>
to select
all
channels. Then press
<MACT>
to reduce this selection again to only those
that are
active
. For each input channel the associated connector number is dis-
played.
Menu
SENSOR PROGRAMMING
:
Display of connector number and channel
Connector: 0 Channel : 00
To select next input channel press
To select previous input channel press
To accept the selection of all possible channels press
<MALL>
To reduce selection to all active channels press
<MACT>
12.3.2 Measuring point designation
Each measuring point can be assigned a 10-character alphanumerical designa-
tion (all ASCII characters) to denote as clearly as possible the type of sensor,
measuring location, and / or purpose. This designation is included in all stan-
dard measured value displays. In an output via the interface the measuring
point designation appears in the program header as ´
DESIGNATION
´ and also
in the measured value list (see Manual 6.6.1).
Input in the ´Designation´ function (see 10.5)
Designation : Temperature
Certain control characters at the beginning of the designation have special
functions :
´*J´ This defines a temperature sensor (Ntc, Pt100) as reference for external
cold junction compensation. (see 9.2.7, Manual 6.7.3).
´#J´ This means that an internal cold junction sensor is to be used for a thermo-
62 ALMEMO
®
5690-2M
* SENSOR PROGRAMMING *
±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±
±
Connector: 0 Channel: 00
Comment:
Temperatur
Averaging mode:
CONT
Locking mode:
5
7 Limit max:
35.0 °C
7 Limit min:
-----
5 Base:
-----
5 Factor:
-----
5 Exponent:
0
4 Zero correct.:
-----
4 Slope correct.:
-----
2 Dimension:
°C
1 Range:
NiCr
MALL M PRINT ESC