11. Measuring with the measuring menus
The quickest way to set the required output format is by pressing
<FORM>
; (for
print layouts see Manual 6.6.1).
Change format
<FORM>
Format, adjacent columns ´
n
´:
Cycle timer
00:02:00 Sn
Change format
<F
ORM>
Format, table ´
t
´:
Cycle timer:
00:02:00St
To start cyclic measuring point scan press
<START>
The following symbols will be displayed as verification in the
status bar
will be displayed continuously
, i.e. so long as the measuring operation is
running.
The start arrow lights up
´
©
´
Lights up when data is being output via the interface
´
COM
´
Lights up when measured values are being saved (see 12.1.2)
´
REC
´
To stop cyclic measuring point scan press
<STOP>
´
ll
´
11.3.3 Memory space, memory output, clearing the memory
When measured values are being recorded the
Memory capacity free
function
continuously displays the memory capacity still available. Selecting this function
enables two soft keys, one for direct memory output and one for memory clear-
ing. The output format is as set in the cycle (see 11.3.2 and 12.1.2)
Function
Memory free
e.g.
Memory free
0378.4
KB
To output the memory (see 12.2.6):
<PRINT>
To clear the memory
<CMEM>
11.3.4 Output menu functions
Each measured value menu, together with all its displayed functions, can be
output via the interface to a printer or computer; (for connection of peripheral
devices, see Manual 5.2). If you have selected the standard display, pressing
the key
<PRINT>
will initiate printout of the following protocol.
Print measured value menu
<PRINT>
Measuring point, measured value, designation
01: +0023.5 °C Temper-
ature
MAXIMUM VALUE : 01:+0020.0
°C
MINIMUM VALUE : 01:-0010.0
°C
PRINT TIMER: 00:01:23
Memory capacity - total / free in KB
MEMORY:S0512.1 F0324.4 A
The protocols for individual functions is listed in Section 6.6.1.
38 ALMEMO
®
5690-2M