Measuring point scan and output
11.3 Measuring point scan and output
Measuring point scanning is used to acquire
measured values from all measuring points ei-
ther manually at certain times or cyclically over
a specified period; these values can be saved
on the computer or output to the printer; (see
Manual 6.5).
This can be performed e.g. in the
Data logger
menu :
11.3.1 Once-only output / saving of all measuring points
Once-only manual measuring point scans for acquiring the current measured
values from all active measuring points
(see Manual 6.5.1.1)
can be initiated by
pressing
<MANU>
. If the time-of-day is required in the display, it must first be
set (see 12.1.1). The output format can be set in the
Cycle timer
function (see
Once-only manual measuring point scan :
<MANU>
In
the
status bar
as verification the following symbols will briefly be displayed
:
The start arrow will light up briefly and then go out again
´
©
´
Lights up when data is being output via the interface
´
COM
´
Lights up when measured values are being saved (see 12.1.2)
´
REC
´
Each time the key is pressed again after this the measured values will be pro-
cessed with the associated measuring time.
11.3.2 Cyclic output / saving of all measuring points
For cyclic recording and output of measured values (see Manual 6.5.1.2) the
cycle and the output format must be programmed accordingly. The measuring
operation can be started by pressing
<START>
and stopped by pressing
<STOP>
. Each time a measuring operation starts, if the device has been so con-
figured, the maximum, minimum, and average values of all measuring points
will be cleared (for default setting see 12.5.8).
So long as no measuring operation has been started the
Cycle timer function
displays the cycle. Once the function has been selected (see 10.4) the cycle
can be entered directly (see 10.5). Once started the timer counts down to the
next cycle.
Function
Cycle timer
:
Cycle timer
00:02:00
S
Cycle (
hh:mm:ss
), Saving ON, List format
ALMEMO
®
5690-2M
37
C © REC COM l© ©l R01 * ´´´´´´µµµµ¶
±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±
Time: 12:34:56 Date:01.01.04
Cycle timer:
00:00:30 nS
Memory free:
508.3 kB
Number:
01-001 A
01: 244.5 °C
NiCr Temperatur » H º
7 Limit max:
250.0 °C
Maximum value:
245.7 °C
7 Limit min:
230.0 °C
Minimum value:
224.1 °C
START MANU M PRINT ESC