8
General Disassemble
186
Agilent U1602A/U1604A User’s and Service Guide
2
Open up the battery cover and unplug the 3- wire connector from PCBA
board using a tweezer.
Figure 8-4
Unplug 3-wire connector from PCBA board using tweezer
To replace battery pack
For battery replacement purpose, you can remove battery pack without
removing stand, follow the steps below:
WA RN I N G
To avoid electrical shock or fire during battery replacement:
•
Disconnect test leads, probes and power supply before opening case
or battery doo
r.
•
Do not operate with battery cover open.
•
Use only insulated probes and leads specified.
•
Only replace batteries with 7.2 V Ni-MH custom pack