9
Characteristics and Specifications
204
Agilent U1602A/U1604A User’s and Service Guide
General Characteristics
Table 9-3
General characteristics
Power Adapter
Line voltage range 50/60 Hz, 100 – 240 VAC
Output voltage 12 VDC
Battery
U1571A Ni-MH rechargeable battery pack 7.2 V
Operating time: 4 hours
Charging time (instrument in “off” condition): Approximately 5 hours
Charging time (instrument in “on” condition): Approximately 14 hours
Allow ambient temperature during charging: 10
o
C to 40
o
C
Dimension (HxWxD)
24.1 cm height x 13.8 cm width x 6.6 cm depth
Weight
1.5 kg
I/O
USB 2.0 Full Speed client (standard) and USB 2.0 Full Speed host
(optional). Firmware upgrade through USB.
Warranty
3 years
Operating Environment
Temperature
Operating full accuracy: 0
o
C to 50
o
C
Non-operating: –20
o
C to 70
o
C
Humidity
Operating full accuracy to 80% RH at 40
o
C
Altitude
Operating: Up to 2000 m
Non-operating: 15000 m (50000 ft)
Shock and Vibration
Tested to IEC/EN 60068-2
ESD tolerance
± 4 kV
Safety Compliance
IEC 61010-1:2001 / EN61010-1:2001
Canada: CSA C22.2 No. 61010-1:2004
USA: UL 61010-1:2004
Pollution degree 2
This instrument is rated for indoor use only.