3
Using the Scope Functions
68
Agilent U1602A/U1604A User’s and Service Guide
This XY mode is useful in comparing the voltage levels of two waveforms
point by point and study the frequency and phase relationships between
the two waveforms. To enable the XY mode, follow the instructions below:
1
Enter the Scope menu by pressing the
button.
2
In Scope menu, press F3 to access Delay submenu.
3
On page 1/2, press F3 softkey to display a pull- up menu for the YT and
XY time base selections.
4
Select YT format for volts- versus- time display or XY format for
volts- versus- volts display.
Figure 3-15
XY format display
Main/Zoom Mode
The Main mode is the normal viewing mode for the instrument. You can
use the Main function to return to the view before magnification, where
the horizontal scale factor would then return to its original level.