1
Getting Started
34
Agilent U1602A/U1604A User’s and Service Guide
To Power-on instrument
✔
To turn the instrument on or off, press and hold
button located
at the lower left corner of the front panel. The instrument loads the
last configuration setup from its non- volatile memory of the instrument.
To Reset to Default Factory Settings
✔
The default factory setting returns the instrument to its known
operating condition. Pressing the F1 Factory Setup softkey removes all
previous user- defined configuration settings of the instrument. Once
deleted, these user- defined settings cannot be recovered.
1
Press and hold
button to enter Save/Recall menu.
2
Select F1 softkey to enter Save/Recall Setup mode.
3
On page 1/4 of Save/Recall mode, press F1 to restore factory setup.
The major default settings are:
Horizontal
Main mode, 100
μ
s/div scale, 0 s delay, center time reference,
window off
Vertical
Vertical channel 1 on, 5 V/div scale, DC coupling, 0 V position,
probe factor to 1X
Trigger
Edge trigger, Auto sweep mode, 0 V level, channel 1 source, DC
coupling, rising edge slope
Display
Vector on, 50% contrast, graticule grid, persistence off
Acquire
Normal mode, Run/Stop to Run, cursors and measurement off
DMM
Voltmeter set to DC, relative off, auto- measurement
Ohmmeter set to Resistance, relative off, auto- measurement
Temperature meter set to
°
C, relative off